Collimating PT Optics

In outline, the PT collimation procedure is:

Step-by-step (not tested yet):

Primary Mirror should be Level During installation the primary mirror is made level in its cell. Confirm that this was done by checking the log and asking the person who most recently reassembled the telescope. Three tip-tilt screws, one under each of the mirror support whiffles, are used to make the mirror's edge a fixed distance from the top of the mirror cell. These screws are never to be adjusted "on the sky" (photo).
Remove Instruments If the instruments are not already off the telescope, follow the instructions for removing the primary mirror up to and including the step "Remove corrector lens." This is a normal state during reassembly after aluminization. NB: At this point, the telescope should be pinned in both axes and the Dec axle tension should be OFF. The HALT MOTORS button on the control panel should be IN.
Add trim weights We want the telescope to be close to balanced. You do not need to be in perfect balance but the telescope should not be too heavy to move safely.

I can imagine in the usual scheme of things the sliding black counterweights will be out of reach at this point. If they're not, bring them down and you might not have to fool with the trim weights.

Attach trim weights to the primary cell until the telescope feels balanced or until your run out of room for weights. Trim weights the arc shaped lead weights that fit on the outer bolt circle of the primary cell. You'll find them in the bottom drawer of the toolbox. Be sure to use the screws in the bottle labeled "Trim Weights." The telescope may still be top heavy after adding all the trim weights. It will certainly want to "fall" north.

Install Collimating Telescope Locate the collimating telescope (it's the same one used on the 2.5 m) and its PT mounting fixture. Mount the telescope on the 20-inch telescope. NB: The screw holes in the back of the primary cell are not tapped through to prevent accidentally contacting the primary mirror. You need screws of exactly the correct length to mount the collimating telescope.
Pin Telescope at Service Position You may want to roll the tool boxes out of the way for this. Position the ladder north of the telescope. With someone holding the telescope tube, pull the pin on the DEC axle and move the top of the telescope north, towards the horizon. Pin the DEC axle near the horizon. I'm not certain if there's a position exactly at the horizon; if not, it's about six degrees below the horizon.
Adjust Sliding Counterweights Move both of the black sliding counterweights to the lowest position, near the telescope center section or to a "balance when horizontal" situation.
Remove Light Shield You probably want to check the secondary mirror centering before removing the light shield, since if it's OK, you don't need to remove the shield.

Remove the light shield around the secondary mirror by unscrewing the four screws around its top edge. It's not possible to remove it from the telescope tube so tie it up around one of the secondary supports. Be careful not to touch the secondary mirror.

Center Secondary With direct viewing through the collimating telescope focus on the center mark on the mirror (you might need to open the dome to get enough light to see the mark). The mark is on a piece of Kapton tape stuck to the mirror's center.

Use the four nylon screws on the side of the secondary mirror cell to center it in the collimating telescope. This is much easier with two people. Be careful not to tighten the screws too tight; they should only be snug enough to keep the mirror from slipping around.

Replace Light Shield When you're satisfied that the centering is OK, attach the light shield.
Adjust Secondary Tilt Turn on the reticle light and focus so you can see the outgoing and return targets clearly. Adjust the screws on the plate holding the secondary cell to center the return image in the collimating telescope, right over the outgoing image. These screws pull the mirror cell up against a spherical washer so you need to loosen some before tightening others. Don't tighten them too much or you'll warp the secondary cell and mirror. Use only enough pressure to keep the mirror in place without straining the cell.
Double-check Secondary Centering Turn off the reticle and refocus to see the centering target directly. It should not have moved but if it did, remove the light shield, re-center, and repeat the tilt adjustment.
Return to Zenith Unpin the telescope, push it to zenith, and replace the pin.
Remove Collimating Telescope Unscrew the collimating telescope from the primary cell. Store the mounting unit in the bottom drawer of the tool box and return the telescope to wherever it's normally stored.
Remove Trim Weights Remove the trim weights from the back of the primary cell, making sure the original weights are in their proper locations. Return the trim weights to the bottom drawer of the toolbox.
Install Corrector Lens After checking that the primary baffle is in place (install it if it isn't), install the top (first) corrector lens.
Install Filter Box Use the pneumatic lift to install the filter box.
Install Eyepiece Clamp the eyepiece holder into the back of the filter box after removing the cover plate ("moth blocker").
Balance the Telescope See Wendell Jordan. Generally speaking, if you've got the trim weights in the right place, balancing requires only that the two black counterweights be positioned correctly. DEC Balance should be checked at zenith and at far south positions. Wendell Jordan is currently the only on-site person trained to balance the PT.
Tension DEC drive See Wendell Jordan. This procedure should be done only by trained personnel to avoid damaging the drive disks. Using the spring scale, tighten the DEC drive tension screw until breakaway occurs at about 10 lbs. Maybe a bit more.
Check Balance Use the drive motor current meters as a sanity check on the balance. Check balance near the zenith and at high zenith distance.
Initialize Telescope Check the telescope coordinate setup and reinitialize at the zenith if necessary.
Bright Star Exam Point to a really bright star and examine the image with an eyepiece. You'll need to open the shutter (toggle switch on the shutter control box) and select the g-band filter (from MOP). You'll need to take the telescope relativly far from its usual CCD focus value to find the eyepiece focus position. Using a high power (15 mm or shorter) eyepiece, study the image. Just out of focus you'll notice some astigmatism. Farther from focus, you'll see a donut of light. Notice the difference inside and outside of focus.

Unless the telescope was incorrectly disassembled and reassembled, the donut of light should be fairly symmetric. If not, you'll need to go to the next step and adjust the primary mirror translation.

Adjust Primary Translation Use the four centering screws around the base of the primary mirror cell to make the donut of light perfectly symmetric. You'll need to loosen one of the screws before tightening its opposite. When you're finished, make sure the screws are snug but not tight. You don't want to pinch the mirror, but you also don't want it sliding around.

Generally speaking, the primary will be well centered if the telescope was disassembled and reassembled according to instructions. You will often not be able to improve on this initial centering but most people cannot resist.

Remove Eyepiece Remove the eyepiece holder and install the filterbox cover (moth blocker).
Install CCD See Jon Brinkmann. Install the CCD dewar on the filter box. Before doing this, you should look closely at the clamp system after removing the cover. Notice the tiny pins poking out of the clamps pointing to the optical axis. These pins center the CCD dewar on the optical axis and are adjusted by setscrews (well, they are the setscrews).

Besides clamping the dewar, you'll also need to connect the cables, dress the CryoTiger hoses, and power up the system.

Adjust CCD Centering Take a few short r-band images near the zenith on a night of good seeing. The stars should look similar across the field, right into the corners of the CCD. Any asymmetries should be radially symmetric, that is, things should point to the center, and more or less similar in size at similar radii.

Image elongation in a large area of the field that cannot be attributed to tracking errors may indicate that the dewar is not centered on the optical axis (the dewar carries the second corrector lens, which needs to be centered on the optical axis to perform well). Consider sliding the CCD dewar along that direction to remove the elongation.

This is done by unclamping the dewar (make sure the latches are in the "safety" position!) and adjusting the centering pins. It's worthwhile doing the trigonometry ahead of time (or take your calculator to the dome) to turn the three screws the right amount to get the direction correct. You want to move the CCD in the direction that makes things symmetric. Typical motions to start with are around 0.5 mm translation. You can figure out the direction but it's just as fast to try one and see what happens.

This test should be done late in the night, well after the telescope has thermally relaxed since the far field images change character quickly and oddly until the telescope has had a chance to stabilize.

HOME



A. Uomoto, August 28, 2001