1). Background.
The Atomic Force Microscope is a scanning probe microscope where an image is collected based on the interaction between the sample and a point source, in this case the tip.
In AFM, a tip (most commonly made of silicon or silicon nitride (Si3N4) approaches the sample to within interatomic distances (approximately 10 Å). The tip (3-15 microns in length) is mounted at the end of a spring cantilever (approximately 100-500 microns in length).
The spring constant (between 0.10-0.34 N/m for Contact-Mode Silicon tips) of such a tip is less than the equivalent spring between atoms of the sample’s. As the tip is rastered across the sample, it fluctuates with respect to the surface features of the sample.
These fluctuations are caused by interactions (electrostatic, magnetic, capillary, Van der Waals) between the tip and the sample. By measuring the displacement of the tip, a topographical image can be generated.
2). Image Creation.
The AFM probe actually
never moves; rather, it is the sample that is moved in the X, Y and Z,
direction by a PZT (piezoelectric) actuator that defines the scan area.
A laser beam is bounced off a prism and focused onto the end of the cantilever,
slightly behind the tip, where it is reflected back to a 4-quadrant photo
diode detector. The cantilever deflects as it responds to atomic force
variations between itself and the sample and the detector measures the
deflection.
The reflected beam is focused such that difference signal between the four sections of the diode is zero. When the cantilever deflects up, more light illuminates the upper half, creating a more positive signal; the converse is true when the cantilever deflects downward.
Because of the detector design, the instrument is sensitive to ambient light and electrical signals, which cause noise in the measurement. Thus, always be sure to cover the instrument with the protective magnetic shield. Additionally, extraneous vibrations from walking and talking may disrupt the integrity of the data collection process. The AFM is stabilized on a vibration isolation table to reduce the problem of vibrations due to movement. Acoustical vibrations are a common problem with no real systematic cure.
3). The Force-Distance Curve
As the atoms of the tip and sample are brought together they initially weakly attract each other. This attractive force increases until the interacting atoms are so close that their electron clouds begin to repel one another electrostatically. This electrostatic repulsion continues to weaken the attractive force as the separation distance decreases. The attractive force goes to zero (i.e., approaches the limit of zero) when the distance comes within the length of a chemical bond (a few Angstroms). Once the total Van der Waals forces (repulsive forces) are positive, the atoms are in “contact”.
Notice that the slope of the Van der Waals forces curve is quite steep in the contact region. As a result, the van der Waals forces dominate almost any other force that attempts to push the atoms closer together. Thus, in AFM, the cantilever bends when it pushes the tip against the sample’s surface, rather than probing it further into the sample
4). Scan Types.
Non-contact Mode –low resolution (more susceptible to noise), contaminants may interfere with oscillation
Lateral Force – records measurements of friction; the degree of torsion of the cantilever is proportional to the surface friction caused by the lateral force exerted on the probe; measured by the left and right deflection of the laser
Force Modulation – records measurements of elasticity or viscosity, particularly useful for distinguishing otherwise smooth surfaces
Tapping Mode – eliminates frictional forces by intermittently contacting the surface and oscillating with sufficient amplitude to prevent it from being trapped in by adhesive forces; less destructive than contact mode; good compromise between contact and non-contact modes
Lift Mode – two-pass technique that measures magnetic and electric forces above the sample surface. On the first pass of the scan the sample’s surface topography is recorded; on the second pass, the tip scans some set distance above the surface and records the force measurements.
1). Open the Advanced True Image Software on your computer’s desktop.
2). Insert the probe mount into the probe holder.
3).
Align the laser beam on the probe.
4). Prepare a sample.
5).
Insert the sample.
6).
Sample Positioning.
7).
Detector Alignment.
8).
Auto Approach.
9).
Image Acquisition.
Atomic Force Microscopy
is considered a wonderful technological advance because it has opened up
the window to imaging non-conducting materials. The Atomic Force Microscope
has a wide range of applications, from contact lens production to semiconductors
to polymers, mostly due to its ability to image at the sub micron level
(<10-6m). A collection of images ranging from biochemical
substances to actual atoms can be found at: http://www.di.com/theater/ut_surf.html.
Similarly, http://www.thermomicro.com/apps/index.html
categorizes
a myriad of excellent AFM images by field.